Skip to main content

Table 4 Reaction flow and reaction parameters

From: Development of a chemiluminescent immunoassay based on magnetic solid phase for quantification of homocysteine in human serum

Reaction step

Contents

parameters

Step 1

Add calibrators, samples, quality controls

30 μL

Step 2

Adding a reducing agent

15 μL

Step 3

Mix well and react in a warm bath

37℃, 5 min

Step 4

Add enzyme

15 μL

Step 5

Magnetized particulate reagent

30 μL

Step 6

Addition of anti-reagents

30 μL

Step 7

Mix well and react in a warm bath

37℃, 30 min

Step 8

Add cleaning solution for chemiluminescence system and mix well.

300 μL

Step 9

Repeat step 8 for magnetic separation of the reactants

3 times

Step 10

Addition of chemiluminescent substrate solution for fully automated immunoassay for luminescence detection.

200 μL